(MQQFC145) Quartz Flow Cell, 10 mm Path, 600 µL, Bottom M5 Thread
$528.00
JGS1 quartz flow cell with a 10 mm optical path, 600 µL nominal capacity and two opposing optical faces around a 3.5 × 17.5 mm window. This version has an M5 thread at the bottom of the body. The bare body is 12.5 × 12.5 × 36 mm, excluding fittings and tubing. Standard 80 construction; price per cell.
Description
MQQFC145 is the 600 µL black-wall flow-cell version with an M5 thread at the bottom of the body. It combines a 10 mm measurement path with a 3.5 × 17.5 mm optical window and two opposing optical faces.
- Path Length
- 10 mm
- Sample Volume (Nominal)
- 600 µL
- Opening Configuration
- Two Fluid Connections with Bottom M5 Thread
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 10 mm |
| Light Transmission Configuration | Two-Way Light |
| Optical Window Dimensions | 3.5 × 17.5 mm |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 600 µL |
| External Dimensions | 12.5 × 12.5 × 36 mm |
| Cell Body Dimensions | 12.5 × 12.5 × 36 mm |
| External Dimensions Scope | Body Only |
| Body Height | 36 mm |
| Base Dimensions | 12.5 × 12.5 mm |
| Material | Quartz |
| Material Grade | JGS1 |
| Port Count | 2 |
| Fabrication Method | Standard 80 |
| Opening Configuration | Two Fluid Connections with Bottom M5 Thread |
| Port Layout | Top connection at the left in the referenced original; Top connection at the right in the referenced original |
| Pack Quantity | 1 Piece |
| Package Contents | 1 × Quartz Flow Cell |
Choose this option
Choose this version when your design calls for the lower M5 thread and the shorter 36 mm bare-body height. The M5 designation belongs to the bottom detail; it does not identify either of the upper fluid-port threads.
What makes this option different
- 10 mm path / 600 µL nominal capacity: these identify the measurement region. Nominal capacity is not a minimum working-fill specification or the total external circuit volume.
- 3.5 × 17.5 mm optical window: check the beam size and position against this opening; the window dimensions do not specify a tube bore.
- Two opposing optical faces: these support transmission through the stated path. The black surround keeps optical access confined to the exposed window regions.
- Bottom M5 thread: check the lower interface separately from the two top fluid connections. A nominal thread code does not establish pitch, a pressure rating or a universal installation role.
- 36 mm bare-body height: the 12.5 × 12.5 mm footprint and body height exclude the projecting fittings and tubing.
Using this cell
Support the body while attaching or removing fittings. Flush the measurement region with a suitable blank, remove bubbles from the illuminated area and keep the same seating and orientation for blank and sample. Keep tubing loads off the optical assembly.
First-use check
- Inspect the optical faces, body and connection edges for damage or residue.
- Check holder support, beam direction and connection clearance before completing the liquid circuit.
- Flush with a suitable blank, remove bubbles and establish a stable reference before measuring the sample.
Fit and optical alignment
The bare body measures 12.5 × 12.5 × 36 mm. Allow separate space for the top fittings and tubing bends. Check holder support, optical beam access and the lower M5 thread as distinct requirements. Check beam height with your holder before ordering; window position alone does not establish compatibility.
View the English selection reference.

Solvents and cleaning
Standard 80: Standard 80 is an adhesive-bonded construction for routine room-temperature aqueous samples and buffers. Avoid organic sample solvents and organic cleaning agents on this bonded assembly. Check exposure for every wetted fitting, seal and tube as well as the cell.
Do not use hydrofluoric acid on quartz. Avoid hot strong alkalis and abrasive cleaning. Wipe exterior optical faces gently with non-abrasive optical tissue, and keep handling loads off projecting connections.
Included with this option
The listed price is per flow cell. Specify the matching fittings and external tubing required by your circuit when ordering; optical-window dimensions do not specify those fluid accessories.
Availability and ordering
Contact us to confirm current availability and dispatch timing for your quantity. For a required delivery date, check timing before checkout; preparation and production are separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz projects involving optical paths, sample regions, optical access, connection geometry and mounting dimensions. Modified designs and prototype batches are quoted separately. Send your drawing, holder details, sample conditions and quantity through the contact page, and include MQQFC145.
Design notes for this configuration: Quartz Cell with a Bottom M5 Feature: Defining the Mating Part
Additional information
| Weight | 3000 g |
|---|
You must be logged in to post a review.