(MQQCU023) Quartz Rectangular Sample Cell, 0.25 mm path, 65 µL nominal capacity, Internally Grooved Chamber
Description
MQQCU023 is a JGS1 quartz rectangular sample cell for UV/Vis spectroscopy with a 0.25 mm optical path and 65 µL nominal capacity. This cell is open at both ends and uses an internally grooved chamber, so the way it is supported, filled and aligned is part of the selection. It suits setups built around this specific open-through geometry rather than a capped bench-style cuvette.
- Path Length
- 0.25 mm
- Sample Volume (Nominal)
- 65 µL
- Opening Configuration
- Open at Both Ends
Selection Specifications
| Product Kind | Specialized Cell |
|---|---|
| Cell Type | Rectangular Sample Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 0.25 mm |
| Light Transmission Configuration | Two-Way Light |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 65 µL |
| External Dimensions | 7 × 2.25 × 52 mm |
| Body Height | 52 mm |
| Base Dimensions | 7 × 2.25 mm |
| Material | Quartz |
| Material Grade | JGS1 |
| Fabrication Method | Molded 83 |
| Opening Configuration | Open at Both Ends |
| Pack Quantity | 1 Piece |
| Package Contents | 1 × Quartz Rectangular Sample Cell |
Choose this option
Choose MQQCU023 when your method needs a very short 0.25 mm transmission path in a slim rectangular body, and when your fixture is made for an open-through cell. The internally grooved chamber is part of the working geometry, so it is best selected when your holder and sample-loading approach are already matched to this format.
What makes this option different
- 0.25 mm path length: suited to methods that require a very short fixed transmission path.
- Open at both ends: support, sealing arrangement and sample handling belong to the fixture, not to an integrated cap or stopper.
- Internally grooved chamber: this is a shaped internal sample region rather than a plain open well.
- Slim body envelope: the 7 × 2.25 × 52 mm dimensions are important when checking holder clearance.
Structure and dimensions

Using this cell
This cell is open at both ends, so it is typically used within a holder or assembly that supports the body and manages sample introduction through the ends. Fill carefully and inspect the measurement region for trapped bubbles, because bubbles or streaking can interfere with a short-path reading.
Keep the same orientation for blank and sample measurements. After filling, remove exterior droplets and handle the optical faces carefully to avoid fingerprints or wipe marks in the light path.
First-use check
- Inspect the optical faces and open ends for chips, residue or handling marks.
- Place the cell in the intended holder and verify that the beam passes through the intended optical region.
- Run a blank in the same orientation you will use for the sample, then refill carefully and recheck for bubbles.
Fit and optical alignment
View the English selection reference. Also available in the product gallery.
The external dimensions are 7 × 2.25 × 52 mm. Check holder clearance across the full body length and confirm how the open ends will be supported in your fixture.
Z Height is not established for this SKU, so beam position should be checked directly against your holder or optical setup. A matching path length alone does not confirm fit.

Solvents and cleaning
Molded 83 is the current fabrication designation for this SKU. This type is commonly chosen for methods that may involve organic solvents or elevated-temperature work, but suitability still depends on the complete wetted assembly and the actual liquid conditions used in your method.
Keep hydrofluoric acid away from quartz. Avoid hot strong alkalis and abrasive cleaning methods. Clean the optical faces gently with a suitable non-abrasive wipe, and avoid scratching the narrow measurement surfaces.
Included with this option
1 Quartz Rectangular Sample Cell.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz production for requirements involving dimensions, optical path length, sample volume, optical windows or Z Height, closures and connections, fabrication and tolerances. Send a dimensioned drawing or labelled sketch and identify the dimensions and tolerances that are critical to your application.
Send a product-selection enquiry with SKU MQQCU023, your holder or drawing details, sample conditions, required quantity and delivery destination.
Design notes for this configuration: 0.25 mm Quartz Groove Cell for Thin Sample Layers
Additional information
| Weight | 50 g |
|---|
You must be logged in to post a review.










Reviews