(MQQFC103) Quartz Flow Cell, 10 mm Path, 2.6 mL, 8 mm OD / 5 mm ID Tubes
Description
MQQFC103 combines a clear rectangular quartz sample body with two opposed tube connections. Its 10 mm optical path is distinct from the 30 mm body length and the 46 mm fixed tube-axis span, making those three measurements important to separate during selection.
- Path Length
- 10 mm
- Sample Volume (Nominal)
- 2.6 mL
- Opening Configuration
- Two Opposed Fluid Tubes
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 10 mm |
| Light Transmission Configuration | Four-Way Light |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 2.6 mL |
| External Dimensions | 46 × 12.5 × 12.5 mm |
| Cell Body Dimensions | 30 × 12.5 × 12.5 mm |
| Uncapped Cell Dimensions | 46 × 12.5 × 12.5 mm |
| External Dimensions Scope | Uncapped Cell |
| Internal Dimensions | 26 × 10 × 10 mm |
| Material | Quartz |
| Material Grade | JGS1 |
| Port Count | 2 |
| Tube Outer Diameter (OD) | 8 mm |
| Tube Inner Diameter (ID) | 5 mm |
| Fabrication Method | Molded 83 |
| Opening Configuration | Two Opposed Fluid Tubes |
| Port Layout | Opposed end A, shown on the left of the exact original; Opposed end B, shown on the right of the exact original |
| Pack Quantity | 1 Piece |
| Package Contents | 1 × Quartz Flow Cell |
Choose this option
Choose this version when the method specifies a 10 mm path and a 2.6 mL nominal sample region, with 8 mm OD / 5 mm ID connections at opposite ends. Check optical alignment and the complete liquid-circuit clearance together.
What makes this option different
- 10 mm optical path: this is the specified measurement distance, rather than the 26 mm internal length, 30 mm body length or 46 mm fixed tube span.
- 2.6 mL nominal capacity: the stated internal region is 26 × 10 × 10 mm. External tubing adds its own volume to the circuit.
- Four optical faces: the clear rectangular body provides optical access around the sample region; it does not imply four independent liquid channels.
- Two opposed tubes, 8 mm OD / 5 mm ID: match the connection dimensions and support attached tubing so it does not load the cell.
- Separate body and assembly dimensions: the bare body is 30 × 12.5 × 12.5 mm; the fixed tube-to-tube span is 46 mm.
Using this cell
Support the body while attaching tubing to the two opposed connections. Flush with a suitable blank, remove bubbles from the optical region and maintain consistent seating and beam alignment between blank and sample.
First-use check
- Inspect the optical faces, body and connection edges for damage or residue.
- Check holder support, beam direction and connection clearance before completing the liquid circuit.
- Flush with a suitable blank, remove bubbles and establish a stable reference before measuring the sample.
Fit and optical alignment
Allow for the 46 mm fixed tube-axis span before adding external tubing and its bend radius. Check support for the 30 × 12.5 × 12.5 mm body, the chosen optical direction and clearance around both tube ends. The fixed span does not include your downstream fittings or hoses.
View the English selection reference.

Solvents and cleaning
Molded 83: Molded 83 identifies the integrated-forming construction mapped for this option. Select sample and cleaning conditions for the complete wetted assembly, including connections and any tubing you add. Construction alone does not establish a pressure rating or compatibility with every solvent.
Do not use hydrofluoric acid on quartz. Avoid hot strong alkalis and abrasive cleaning. Wipe exterior optical faces gently with non-abrasive optical tissue, and keep handling loads off projecting connections.
Included with this option
One quartz flow cell with the two fixed opposed tubes shown. The listed price is per cell; external tubing and a holder are not stated as included.
Availability and ordering
Contact us to confirm current availability and dispatch timing for your quantity. For a required delivery date, check timing before checkout; preparation and production are separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz projects involving optical paths, sample regions, optical access, connection geometry and mounting dimensions. Modified designs and prototype batches are quoted separately. Send your drawing, holder details, sample conditions and quantity through the contact page, and include MQQFC103.
Design notes for this configuration: Large-Port Quartz Flow Cell: Planning the Tubing Route (MQQFC103)
Additional information
| Weight | 50 g |
|---|
You must be logged in to post a review.