(MQQFC130) Quartz Flow Cell, 3 mm Path, 300 µL, Same-Side Tubes
Description
MQQFC130 is a JGS1 quartz flow cell for UV/Vis setups that need a short 3 mm optical path in a compact inline format. Its 300 µL nominal capacity, square body and two side tubes on one face make it easier to plan the measurement zone and the tubing route together. The six-sided optical access layout also leaves optical access around the sample region from multiple directions.
- Path Length
- 3 mm
- Sample Volume (Nominal)
- 300 µL
- Opening Configuration
- Two Side Tubes
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 3 mm |
| Light Transmission Configuration | Six-Way Light |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 300 µL |
| Material | Quartz |
| Material Grade | JGS1 |
| Port Count | 2 |
| Tube Outer Diameter (OD) | 4 mm |
| Tube Inner Diameter (ID) | 2 mm |
| Fabrication Method | Molded 83 |
| Opening Configuration | Two Side Tubes |
| Port Layout | Upper side tube on the left face as drawn; Lower side tube on the same left face as drawn |
| Wall Thickness | 1.5 mm |
| Pack Quantity | 1 Piece |
| Package Contents | 1 × Quartz Flow Cell |
Choose this option
Choose MQQFC130 when you need a small quartz flow cell with a defined 3 mm path and side connections that stay grouped on one side of the body. This layout is useful when holder space is limited or when the tubing approach should not block the optical area around the cell.
What makes this option different
- 3 mm optical path with a 300 µL nominal capacity for compact UV/Vis flow measurements.
- six-sided optical access layout for optical access around multiple sides of the square sample region.
- Two side tubes on one face so both fluid connections approach from the same side of the cell.
- 4 mm OD / 2 mm ID tube dimensions for planning tubing and connector fit.
Using this cell
The main sample section is square, while the side connections are round tubes. The 1.5 mm wall thickness applies to the main section; it does not describe the 4 mm OD / 2 mm ID side tubes. Six-sided optical access is distinct from the two fluid connections and does not imply six liquid circuits.
Mount the cell so the beam follows the 3 mm optical path through the filled sample region. Connect the liquid circuit with enough support that the body, not the tubes alone, carries the handling load.
Fill gradually and clear bubbles before measurement, especially if the installed tubing layout can trap air near the flow channel. The stated 300 µL is a nominal capacity for the cell and should not be treated as a minimum working volume.
First-use check
- Inspect the optical faces, tube openings and body corners for chips, residue or handling marks.
- Check that your tubing can connect without forcing the cell out of position.
- Run a blank after filling and again after any trapped bubbles have been removed.
Fit and optical alignment
The source geometry shows a square body with 6 mm outer width and 3 mm inner width, plus 1.5 mm wall thickness in the stated body section. The body axis is shown as 43 mm, and each side tube projects 12 mm, so installed clearance needs to include the tube extensions as well as the optical body.
Z Height is not stated for this SKU. If your holder or instrument depends on a fixed beam center height, check that point before ordering rather than relying on the body outline alone.
View the English selection reference.

Solvents and cleaning
This JGS1 cell uses Molded 83 fabrication. Check sample and cleaning conditions for the complete assembly, including any connected tubing and adapters. For organic solvents or heated cleaning, confirm suitability for the actual liquid, concentration, temperature and contact time; the fabrication designation alone is not a complete-assembly limit.
Do not use hydrofluoric acid with quartz. Avoid hot strong alkalis and abrasive cleaning methods, and wipe the optical faces gently with non-abrasive optical tissue.
Included with this option
1 quartz flow cell. The listed price is for one piece.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
If you need a different path length, channel size, body format or tube arrangement, send your drawing or holder details. Custom projects are reviewed separately from this listed SKU, with final specifications, quantity, price and timing confirmed for the project.
Additional information
| Weight | 75 g |
|---|
You must be logged in to post a review.








Reviews
There are no reviews yet.