Quartz 1mm Cuvette With Tubing, 350ul, Molded 83, Two-way Light, 1pc/ea, C012WS
Description
The C012WS is a quartz cuvette with tubing with 1mm path length and two-way light optical configuration — two opposite vertical faces are polished to optical grade, forming the standard transmission geometry for UV-Vis absorbance measurements. With a 350ul internal volume (10 × 1 × 94 mm chamber), the cell is compatible with most commercial spectrophotometers. Fabricated using Molded 83 (>83% transmittance) — precision-molded from a single piece of quartz, eliminating adhesive joints entirely. Features a distinctive long-neck tube design with a 95mm total height, where a cylindrical glass tube (8mm OD, 6mm ID) extends from the compact 1mm cuvette body, enabling connection to external tubing systems for continuous flow measurements.
| Technical Specifications — Cuvette With Tubing | |||
|---|---|---|---|
| Material | Quartz | SKU | C012WS |
| Path Length | 1mm | Internal Volume | 350ul |
| Optical Windows | 2 (Two-Way Light) | Fabrication | Molded 83 |
| Transmittance | >83% | Wavelength | 190–2500 nm |
| Exterior Dimensions | 12.5 × 3.5 × 95 mm | Interior Dimensions | 10 × 1 × 94 mm |
| Closure | None (Open Top) | Package | 1pc/ea |
Spectroscopy Applications
Why Two-Way Light?
Two-way light cuvettes have two opposite vertical faces polished to optical grade, forming the standard transmission geometry for UV-Vis absorbance measurements. The remaining two faces are frosted to minimize stray-light interference, providing clean baseline performance for quantitative spectrophotometry.
Why Molded 83?
Molded 83 uses monolithic fused-quartz construction — no adhesive joints — delivering >83% UV transmittance from 200–2500 nm. Because there are no glue seams, the cell withstands all common laboratory solvents, concentrated acids, and bases (avoid HF). This fabrication is ideal for demanding analytical workflows where chemical resistance, long service life, and maximum optical clarity are priorities.
- Rinse with a compatible solvent (e.g., ethanol or the measurement solvent) before first use to remove any residual particles.
- Do not use abrasive materials or brushes on the polished optical faces — wipe gently with lens tissue.
- Withstands all common laboratory solvents, concentrated acids, and bases. Avoid HF (hydrofluoric acid).
- Store in the original protective case to prevent scratches on the optical windows.
- For baseline correction, fill with your reference solvent, cap the cuvette, and run a blank before sample measurements.
Need a non-standard path length, volume, window configuration, or closure type? We manufacture quartz cuvettes to your exact specifications. Request a custom quote.
Additional information