(C102SS19) Quartz Micro Cuvette, 10 mm Path, 1.75 mL
Original price was: $72.00.$65.00Current price is: $65.00.
C102SS19 is a JGS1 quartz micro cuvette for UV/Vis spectroscopy with a 10 mm optical path, 1.75 mL nominal capacity and Two-Way Light transmission layout. Its 5 mm sample-region width offers the broadest filled measurement zone in this group while keeping the same compact square micro-cuvette body. Supplied as 1 cuvette per price unit, with a stopper included.
Description
C102SS19 is a quartz micro cuvette for UV/Vis work when you need a 10 mm optical path and the largest nominal capacity within this compact square micro-cell format. This JGS1 version combines a 1.75 mL nominal capacity, a 5 mm sample region and a stopper-equipped top, making it a practical choice for straight-through measurements that benefit from a broader filled region in a small cuvette body.
- Path Length
- 10 mm
- Sample Volume (Nominal)
- 1.75 mL
- Closure Type
- Stopper
Selection Specifications
| Product Kind | Cuvette |
|---|---|
| Cell Type | Micro Cuvette |
| Measurement Mode | UV/Vis Spectroscopy |
| Pack Quantity | 1 Piece |
| Path Length | 10 mm |
| Light Transmission Configuration | Two-Way Light |
| Optical Window Dimensions | 5 mm |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 1.75 mL |
| External Dimensions | 12.5 × 12.5 × 45 mm |
| Base Dimensions | 12.5 × 12.5 mm |
| Material | Quartz |
| Material Grade | JGS1 |
| Closure Type | Stopper |
| Closure Included | Yes |
| Fabrication Method | Standard 80 |
| Detection Channel Width | 5 mm |
Choose this option
Choose C102SS19 when your method needs a 10 mm transmission path and you want a broader sample region than the narrower-channel options in the same family. The 1.75 mL nominal capacity and 5 mm sample-region width help when beam footprint and fill coverage are easier to manage in a wider filled zone, while the compact square base still suits micro-cuvette holder formats.
What makes this option different
- 10 mm optical path for standard UV/Vis transmission measurements.
- 1.75 mL nominal capacity in a compact micro cuvette format.
- 5 mm stated sample-region width for a broader filled measurement zone.
- Two-Way Light layout with a stopper included for covered sample handling.
Using this cell
Fill the cell so the illuminated region remains fully covered during measurement, and keep bubbles, droplets and residue off the optical faces. Use a consistent orientation for blank and sample runs to keep the light path and beam position repeatable.
The 1.75 mL figure is the nominal cell capacity, not a minimum working volume. If you plan to work below nominal fill, check in your own setup that the beam still passes fully through the liquid column.
First-use check
- Inspect the optical faces for dust, chips or residue.
- Rinse with a suitable liquid for your method before first use.
- Confirm that the cuvette seats stably in the holder.
- Run a blank and verify that the beam passes through a fully filled, bubble-free sample region.
Fit and optical alignment
The stated external dimensions are 12.5 × 12.5 × 45 mm, with a 12.5 × 12.5 mm base. This helps you check holder fit and top clearance in your cuvette compartment.
Z Height is not stated for this exact option, so fixed-beam instruments should be checked for beam-center alignment before ordering. This is a Two-Way Light cell for straight optical transmission, and the stated 5 mm window/sample-region width should be matched against your beam footprint rather than treated as a certified clear aperture.
Solvents and cleaning
This cuvette uses Standard 80 adhesive-bonded quartz construction. It is generally suited to routine room-temperature water-based samples and buffers.
Avoid organic samples and organic cleaning solvents with this adhesive-bonded cell. Do not use hydrofluoric acid on quartz. Avoid hot strong alkalis and abrasive cleaning, and wipe the optical faces gently with non-abrasive optical tissue or similar laboratory materials.
Included with this option
One cuvette is supplied per sale pack. A stopper is included with this cell.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz projects for changes to dimensions, optical path length, sample volume, optical windows, Z Height, closures and fabrication requirements. Send a dimensioned drawing or labelled sketch together with your instrument or holder details, sample conditions and quantity requirements. Discuss your requirements and quote part number C102SS19.
Additional information
| Weight | 30 g |
|---|
You must be logged in to post a review.






Reviews
There are no reviews yet.