(MQQMC024) Quartz Flow Cell, 10 mm path, 32 µL nominal capacity, 2 mm Optical Opening
Description
MQQMC024 is a JGS1 quartz flow cell with a 10 mm optical path, 32 µL nominal capacity and two top fluid connections. It suits UV/Vis setups that need the familiar 10 mm path in a compact flow format, and the stated 8.5 mm Z Height gives a direct reference for checking beam position in a holder or instrument.
- Path Length
- 10 mm
- Sample Volume (Nominal)
- 32 µL
- Opening Configuration
- Two Top Fluid Connections
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 10 mm |
| Light Transmission Configuration | Two-Way Light |
| Z Height | 8.5 mm |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 32 µL |
| External Dimensions | 12.5 × 12.5 × 35 mm |
| Body Height | 35 mm |
| Base Dimensions | 12.5 × 12.5 mm |
| Material | Quartz |
| Material Grade | JGS1 |
| Port Count | 2 |
| Fabrication Method | Standard 80 |
| Opening Configuration | Two Top Fluid Connections |
| Fitting Material | Stainless Steel |
| Port Layout | Top connection at the left in the referenced original; Top connection at the right in the referenced original |
| Pack Quantity | 1 Piece |
| Package Contents | 1 × Quartz Flow Cell |
Choose this option
Choose this cell when you want a 10 mm flow path in a compact body and prefer tubing access from above the cell. Its 32 µL nominal capacity and 8.5 mm Z Height are especially useful when you need to plan both a small internal volume and a defined beam-center position without moving to a larger body.
What makes this option different
- 10 mm optical path for UV/Vis methods built around the standard cuvette transmission length.
- 32 µL nominal capacity in a compact flow-cell body.
- 8.5 mm Z Height to help check beam-center alignment against your holder or instrument.
- Two top fluid connections with stainless steel fittings for above-body tubing access.
Structure and dimensions

Using this cell
Mount the cell so the optical faces sit correctly in the beam, then connect the fluid lines from the top with the body fully supported. Fill gradually and watch the sample region for trapped bubbles before taking a blank or sample reading.
The stated 32 µL value is the nominal cell volume, not a minimum working volume. In practice, your operating fill should be enough to cover the optical region under the actual flow conditions in your system.
First-use check
- Inspect the optical faces, body edges and fitted connection points before use.
- Verify holder fit and beam alignment with a suitable blank.
- After connection and filling, confirm that the optical region is bubble-free and that the exterior faces are clean and dry.
Fit and optical alignment
View the English selection reference. Also available in the product gallery.
The bare body dimensions are 12.5 × 12.5 × 35 mm. Check holder depth and surrounding clearance separately from the added space required for the top fittings, tubing and bend radius.
The 8.5 mm Z Height is useful for matching the beam center, but instrument fit still depends on your holder, clamp and connection clearance. The fittings are stainless steel, but the exact fitting grade and pressure limit are not stated for this SKU.

Solvents and cleaning
Standard 80 is an adhesive-bonded quartz construction generally used for room-temperature water-based samples and buffer solutions. Avoid organic samples and organic cleaning solvents with this construction.
Do not use hydrofluoric acid with quartz. Avoid hot strong alkalis and abrasive cleaners, and clean the optical faces gently with suitable non-abrasive lab wipes or optical tissue. For the full assembly, also consider the suitability of the stainless steel fittings for your process fluid and cleaning routine.
Included with this option
1 quartz flow cell, sold as 1 piece.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz projects for changes to path length, Z Height, body size, volume, connection layout and related tolerances. Send a drawing or labelled sketch together with holder details, sample conditions and the dimensions that matter most in your setup through the contact page, and mention SKU MQQMC024.
Additional information
| Weight | 50 g |
|---|
You must be logged in to post a review.












Reviews