(MQQFC122) Quartz Counting Flow Cell, 0.5 mm path, 40 µL nominal capacity, In-Line Tube Connections
Description
MQQFC122 is a JGS1 quartz counting flow cell for UV/Vis use, combining a 0.5 mm optical path with a 40 µL nominal capacity in a compact rectangular body. Its two opposed tube connections make it a practical choice when you want a short-path inline cell rather than a bare recessed body that depends on a separate sealing fixture.
The small body helps when space around the optical position is limited, while the stated tube size gives you a direct reference for planning the connected liquid path. This keeps the selection focused on three key points: short optical path, small nominal capacity and straight-through tube access across the cell.
- Path Length
- 0.5 mm
- Sample Volume (Nominal)
- 40 µL
- Opening Configuration
- Two Opposed Fluid Tubes
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Counting Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 0.5 mm |
| Light Transmission Configuration | Two-Way Light |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 40 µL |
| External Dimensions | 30 × 15 × 4 mm |
| External Dimensions Scope | Body Only |
| Material | Quartz |
| Material Grade | JGS1 |
| Port Count | 2 |
| Tube Outer Diameter (OD) | 1.5 mm |
| Tube Inner Diameter (ID) | 1 mm |
| Fabrication Method | Standard 80 |
| Opening Configuration | Two Opposed Fluid Tubes |
| Port Layout | End/side connection at the left in the referenced original; Opposite end/side connection at the right in the referenced original |
| Pack Quantity | 1 Piece |
| Package Contents | 1 × Quartz Counting Flow Cell |
Choose this option
Choose MQQFC122 when your method needs a 0.5 mm path in a compact flow-cell body with direct tube connections on opposite sides. It suits installations where the holder space is limited but the sample still needs to move through a fixed inline route instead of being managed by an external recess-and-gasket arrangement.
What makes this option different
- 0.5 mm optical path for methods built around a short light path.
- 40 µL nominal capacity in a compact quartz flow-cell body.
- Two opposed tube connections for a straight-through inline connection layout.
- 1.5 mm OD and 1 mm ID tube size for planning the connection to your tubing system.
- 30 × 15 × 4 mm body-only dimensions with the projecting tubes outside that stated body size.
Structure and dimensions

Using this cell
Support the quartz body during installation and connect tubing without using the projecting tubes as handles or levers. Fill the cell slowly so the sample reaches the optical region without trapping bubbles, then establish your blank under the same circuit conditions used for measurement.
The 40 µL figure is the nominal cell capacity, not a stated minimum working volume. If your method is sensitive to fill level or bubble position, check that the illuminated region remains fully covered in your actual flow setup.
First-use check
- Inspect the optical faces and tube openings for chips, contamination or handling residue.
- Test holder seating and beam alignment with a suitable blank before running samples.
- After filling, check the optical region for bubbles and repeat the blank in the same installed position.
Fit and optical alignment
View the English selection reference.
The stated 30 × 15 × 4 mm dimensions describe the body only. Leave additional clearance for the tube projections, tubing bends and any fittings used in your assembly.
Z Height is not stated for this model, so beam position should be checked against your holder or instrument before ordering. Body size alone does not establish instrument compatibility.

Solvents and cleaning
This cell is made in Standard 80 adhesive-bonded quartz construction. It is commonly selected for routine room-temperature aqueous samples and buffers.
Avoid organic samples and organic cleaning solvents on this bonded assembly. Do not use hydrofluoric acid with quartz. Avoid hot strong alkalis and abrasive cleaning materials, and clean the exterior optical faces gently with non-abrasive optical tissue.
Included with this option
1 Quartz Counting Flow Cell.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz production for changes to path length, body dimensions, capacity, openings, tube connections and fabrication requirements. Send your drawing or holder details, sample conditions, quantity and delivery destination to discuss your requirements.
Design notes for this configuration: 40 µL Quartz Counting Cell: Bringing a Channel to the Optics (MQQFC122)
Additional information
| Weight | 50 g |
|---|
You must be logged in to post a review.















Reviews
There are no reviews yet.