Demountable Quartz Cell: Loading Powders Through an Open Recess (MQQMC081)
Custom design guide · MQQMC081
MQQMC081 opens into a grooved main plate and a separate cover. Its 8 × 10 mm recess is 0.8 mm deep, giving direct access to the sample area before the cover is placed. For powders or small solid specimens, that access is the main reason to consider this layout.
The situations and proposed changes below are planning examples, not completed customer projects.
Structure and use · Current specifications · Selection details · Proposed changes · Before ordering · Buy or enquire
Reach the sample area before closing the cell
Loading a powder through a narrow neck can make it awkward to reach or clear the measurement area. A removable plate lets you approach the recess directly. It also changes the task: the sample must stay where it is placed when the cover and holder are brought into position.
For a proposed loading trial, place a representative sample in the exposed recess, establish the intended sample surface, and bring the cover into position before locating the assembly in its holder. Observe whether that movement redistributes the material and whether the sample can be removed afterwards. The recess defines the space; sample form and operating orientation determine how it is occupied.

Current MQQMC081 configuration
Open engineering reference R01 and PDF
Source-based dimensions and configuration for discussion. Unconfirmed geometry and tolerances remain to be agreed; this is not a manufacturing-release drawing.

View full-size engineering reference · Download the complete English engineering reference PDF
Make the cover position repeatable
The main plate is 12 × 2 × 45 mm; the separate cover is 12 × 1.25 × 45 mm. Show which edges the holder supports and how the cover is retained. Those two component dimensions do not establish the clearance of a finished mounted assembly.
For powder measurements, use the same loading and surface-preparation method when comparing specimens. The stated 64 µL describes nominal recess capacity, not a powder mass or a guaranteed working quantity. The holder must also leave the illuminated area clear while retaining the cover; cover placement and optical access are part of the same locating problem.

Move the recess while keeping its sample area
Suppose the 8 × 10 mm sample area suits the task, but it sits above the instrument beam. A useful enquiry would keep the target area and 0.8 mm path, then mark a proposed recess centre from a top or bottom datum.
Include the holder contact edges and the cover position on the same sketch. We can then assess edge margins, support and construction together. This describes a new design request, rather than machining an existing finished cell.

Checks to agree before ordering
- Loading: Try placing, covering and removing a representative sample in the intended orientation.
- Location: Define the recess datum, cover retention and holder contact edges.
- Inspection: Specify the critical recess dimensions, tolerances and any optical records needed in the quotation.
Sample and cleaning conditions
Check sample and cleaning liquids against the complete cell, including its bonded seams and any wetted cap or connection. Supply chemical identity, concentration, temperature and contact time. Avoid hydrofluoric acid; strong alkalis, heat and prolonged soaking need specific review. Adhesive-bonded quartz does not provide blanket approval or rejection of organic solvents.
Order this configuration or discuss a change
View MQQMC081 specifications and current pricing. If you need a delivery date, ask about availability and preparation or production time before ordering.
For a modified version, send the brief below with a dimensioned drawing or labelled sketch. We review the proposed configuration and confirm feasibility, quantity, inspection scope, pricing and production time in the quotation.
Copy this custom-enquiry checklist
Reference SKU: MQQMC081
Keep: 8 × 10 mm recess and 0.8 mm target path; proposed recess position: ____
Sample form, loading/removal method and operating orientation: ____
Holder contacts, cover retention, beam direction and centre from datum: ____
Sample and cleaning liquids, concentrations, temperature and contact time: ____
Wavelength, critical tolerances and required inspection/documents: ____
Quantity, required date and destination: ____
Send your sketch and requirements. Mark the features to keep and the changes to assess.
Configuration reference
Configuration reference: MQQMC081 product specifications. Specifications checked 26 October 2026. General principles guide selection; they do not establish measured performance of this cell.