(C104WS7) Quartz Flow Cell, 10 mm Path, 3.5 mL
Original price was: $270.00.$243.00Current price is: $243.00.
C104WS7 is a JGS1 quartz flow cell for UV/Vis work with a 10 mm optical path, Four-Way Light access and 3.5 mL nominal capacity. This version uses two fixed end tubes in 5 mm OD / 3 mm ID quartz, so the connection size is built into the cell rather than added later. Sold as 1 cell.
Description
C104WS7 is a JGS1 quartz flow cell for UV/Vis spectroscopy with a 10 mm optical path and 3.5 mL nominal capacity. Its Four-Way Light layout and fixed 5 mm OD / 3 mm ID tube connections make it a practical choice when you need to plan the optical body, tubing run and mounting space as one assembly.
- Path Length
- 10 mm
- Sample Volume (Nominal)
- 3.5 mL
- Opening Configuration
- Two Fixed Tube Connections
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Pack Quantity | 1 Piece |
| Path Length | 10 mm |
| Light Transmission Configuration | Four-Way Light |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 3.5 mL |
| External Dimensions | 12.5 × 12.5 × 38 mm |
| Cell Body Dimensions | 38 × 12.5 × 12.5 mm |
| Body Height | 38 mm |
| Base Dimensions | 12.5 × 12.5 mm |
| Material | Quartz |
| Material Grade | JGS1 |
| Port Count | 2 |
| Tube Outer Diameter (OD) | 5 mm |
| Tube Inner Diameter (ID) | 3 mm |
| Fabrication Method | Molded 83 |
| Opening Configuration | Two Fixed Tube Connections |
Choose this option
Choose C104WS7 when your method uses a 10 mm flow-through path and you want fixed straight-through tube connections at both ends of the cell. The 5 mm OD / 3 mm ID tube size is the key selection point here, because it affects how the cell joins the rest of your liquid path and how much clearance you need around the installed body.
What makes this option different
- 10 mm optical path in a quartz flow cell for UV/Vis measurements.
- Four-Way Light configuration for optical access from multiple sides of the cell body.
- Two fixed opposed tube connections, each with 5 mm OD and 3 mm ID.
- Compact 38 × 12.5 × 12.5 mm body dimensions, with the tube span extending beyond the body.
Using this cell
Install the cell as part of a suitable liquid circuit and support the assembly so the fixed quartz tubes are not carrying unnecessary bending stress. Fill the cell gradually and watch the illuminated region as liquid enters, so bubbles or droplets do not remain in the measurement zone.
The stated 3.5 mL is the nominal volume of the cell itself. It does not describe the extra liquid held in attached tubing, connectors or any upstream and downstream components.
First-use check
- Inspect the quartz body and both tube ends for chips, cracks or handling damage.
- Test the physical fit in the holder before connecting the full liquid path.
- Fill slowly with a suitable clean liquid and confirm that the optical region is free of bubbles and exterior residue.
- Run your blank after the flow path is fully filled and stable.
Fit and optical alignment
The quoted 38 × 12.5 × 12.5 mm dimensions apply to the optical body. The full installed length is greater because the fixed tubes project from both ends, so check clearance for tubing approach, bend radius and any holder features near the connection axis.
Four-Way Light describes the supported optical access category around the cell body. Z Height is not stated for this model, so body size alone should not be used as a beam-height substitute when holder position is critical.
Solvents and cleaning
This cell uses Molded 83 fabrication. It is commonly chosen for methods that may involve organic solvents or temperature-controlled use, but the suitable routine still depends on the complete wetted assembly, including any connected tubing and fittings.
Do not use hydrofluoric acid on quartz. Avoid hot strong alkalis and abrasive cleaning methods, and wipe the exterior optical faces gently with non-abrasive optical tissue.
Included with this option
Sold as 1 cell.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz flow cells for changes involving dimensions, optical path length, sample volume, optical windows or Z Height, connections, fabrication and tolerances. Send a dimensioned drawing or labelled sketch through the contact page and quote part number C104WS7.
Additional information
| Weight | 30 g |
|---|