(C052WS) Quartz Flow Cell, 5 mm Path, 300 µL
Description
C052WS is a quartz flow cell for UV/Vis measurements with a 5 mm optical path and 300 µL nominal capacity in a compact body. Its Two-Way Light layout and two top M6 fluid connections give you a defined in-line cell format that is easy to place in a holder while keeping tubing above the optical body. This makes it a practical choice when you need a longer path than the 2 mm version but want the same upright connection style.
- Path Length
- 5 mm
- Sample Volume (Nominal)
- 300 µL
- Opening Configuration
- Two Top Fluid Connections
Selection Specifications
| Product Kind | Flow Cell |
|---|---|
| Cell Type | Flow Cell |
| Measurement Mode | UV/Vis Spectroscopy |
| Path Length | 5 mm |
| Light Transmission Configuration | Two-Way Light |
| Optical Window Dimensions | 17.5 × 3.5 mm |
| Wavelength Range — Lower Limit | 200 nm |
| Wavelength Range — Upper Limit | 2500 nm |
| Sample Volume (Nominal) | 300 µL |
| Cell Body Dimensions | 12.5 × 12.5 × 35 mm |
| Body Height | 35 mm |
| Base Dimensions | 12.5 × 12.5 mm |
| Material | Quartz |
| Port Count | 2 |
| Port Thread | M6 |
| Fabrication Method | Standard 80 |
| Opening Configuration | Two Top Fluid Connections |
| Fitting Material | PTFE |
Choose this option
Choose C052WS when your method calls for a 5 mm transmission path and you want a compact flow cell with both connections on the top of the body. The 300 µL nominal cell volume and 12.5 × 12.5 × 35 mm body help when you need to plan optical clearance and fluid routing together without side-entry tubing around the measurement faces.
What makes this option different
- 5 mm optical path for UV/Vis transmission methods built around this path length.
- 300 µL nominal capacity in the cell body, with system tubing volume planned separately.
- Two top M6 ports with PTFE fittings for an upright connection layout.
- Compact 12.5 × 12.5 × 35 mm body with Two-Way Light transmission through opposing optical faces.
Using this cell
Connect tubing to the two top ports and fill the cell slowly so the illuminated region wets evenly before measurement. Before collecting data, inspect the sample region and window faces for bubbles, droplets or residue, then establish your blank after the flow path is fully filled.
The stated 300 µL value is the nominal capacity of the cell itself. It does not include tubing, fittings or other upstream and downstream parts, so total priming and rinse volume should be worked out from the full connected system.
No pressure limit is stated here. If the cell will be used in a pressurized circuit, review the complete connection arrangement and operating conditions before use.
First-use check
- Inspect the quartz body and optical faces for chips, scratches, residue or contamination.
- Check that your holder can seat the body and leave clearance for top-mounted tubing.
- Fill gradually and confirm that the illuminated sample region is free of trapped bubbles before measurement.
Fit and optical alignment
The optical window area is 17.5 × 3.5 mm, and the body dimensions are 12.5 × 12.5 × 35 mm. Use those body dimensions for holder seating, but allow extra space above the cell for the fitted hardware and attached tubing because the stated size does not include projecting connection parts.
Z Height is not listed for this model, so beam position should be checked against your holder or instrument before ordering where vertical alignment matters. The flow connections are both on the top of the cell body; inlet and outlet assignment should be chosen in your own assembly plan.
Solvents and cleaning
This cell uses adhesive-bonded quartz construction and is listed under Standard 80. It is generally selected for routine water-based samples and buffers at room temperature. Avoid organic samples and organic cleaning solvents with this adhesive-bonded construction.
Do not use hydrofluoric acid on quartz. Avoid hot strong alkalis and abrasive cleaning methods. Clean the optical faces gently with non-abrasive optical tissue or another suitable non-scratching method, and check the compatibility of the rest of the wetted flow path as a complete assembly.
Included with this option
The listed price is for one C052WS flow cell.
Additional accessory contents should be confirmed for your order.
Availability and ordering
Order this specification directly, or contact us to check current availability and dispatch timing. If temporarily unavailable, additional preparation or production time may be needed. For a required delivery date, contact us before checkout; preparation or production is separate from carrier transit and customs clearance.
Discuss a custom version
We coordinate custom JGS1 quartz flow-cell projects for changes involving path length, sample volume, optical windows, external envelope and connection details. Send a dimensioned drawing or labelled sketch together with holder information, application wavelength, liquid conditions and target quantity. Discuss your requirements and quote part number C052WS.
Additional information
| Weight | 30 g |
|---|